Robotic ultrasound calibration achieved without external trackers

Tracker-Free Robotic Ultrasound Calibration with a Spherical-Marker Phantom and Threshold-Free Center Localization

Robotics

Summary

Robotic ultrasound machines need to be precisely calibrated so the images they produce match the robot's position. Existing methods require complicated setups or external devices to track calibration markers. This work presents a simpler approach using a special spherical marker and a new image processing method that automatically finds the marker's center without tweaking settings by hand. The authors tested their system and showed it achieves similar accuracy with just one marker as methods needing multiple markers or trackers.

What this means in practice

  • For robotics engineers: Calibrate ultrasound robots accurately without using external tracking equipment, simplifying setup and improving robustness.
  • For medical device manufacturers: Streamline design and production of robotic ultrasound systems by integrating automatic calibration with spherical markers.$Commercial implications: Enables selling robotic ultrasound devices that avoid costly external trackers and complex phantom setups.

Authors

Kyoungmo Koo, Guangshen Ma, Xueding Wang, Mark Draelos

Abstract

Robotic ultrasound (US) calibration is essential for accurately relating US images to the robot coordinate system, but accurate and automated calibration remains challenging because existing methods often require complex phantoms or external 3D trackers. In this work, we develop a tracker-free robotic US calibration framework using a spherical-marker phantom and a highly automated perception pipeline for sphere-center localization. The proposed threshold-free image-processing method localizes the spherical feature based on each US image's intensity distribution, eliminating hand-tuned intensity thresholds and improving robustness across imaging settings without per-system retuning. Multi-pose observations of the spherical fiducial are then used to estimate the US-to-EE transformation without external tracking or prior localization of the sphere center in the robot base frame. We validate the proposed framework on a robotic US platform through repeated sphere scans and further assess the calibrated system using geometrically distinct phantoms with known CAD models. Across 12 cross-validation folds, single-marker calibration achieved sphere-center accuracy and precision of $1.75\pm0.51$ mm and $0.85\pm0.11$ mm, respectively, compared with $1.72\pm0.51$ mm and $0.84\pm0.12$ mm for three-marker calibration. Across the three reconstruction sets, the single-marker calibration yielded pooled post-registration point-to-surface MAE$\pm$SD values of $0.41\pm0.35$ mm for the cone and $0.50\pm0.41$ mm for the triangular prism, closely matching the three-marker results of $0.40\pm0.35$ mm and $0.48\pm0.40$ mm, respectively.